Dräger 8000 NC Instructions For Use Manual page 27

Incubator
Hide thumbs Also See for 8000 NC:
Table of Contents

Advertisement

Operation
Using skin temperature control
Alarms
Alarms
For
deviations greater than ±0.5 °C
between set value
and actual value of skin temperature:
1
Red
Alarm
LED and
2
Red ±0.5
°C
LED flash,
3
display flashes and intermittent tone commences.
The intermittent tone can be suppressed for 10 minutes:
4
Press ® key.
5
Yellow $ LED and red
Alarm
LED are lit.
When measured value is within ±0.5 °C of the set value
again,
2
±0.5 °C LED and
1
red
Alarm
LED go out.
intermittent tone ceases.
5
Yellow
(&.
LED goes out.
If the
sensor plug is disconnected
or
sensor is faulty:
6
Display of 3 dashes in centre of display.
After 15 seconds:
Intermittent tone commences.
6
Display of 3 flashing dashes in centre of display.
7
Red
Alarm
LED and
8
red
Sensor
alarm
LED
are lit.
Then:
• Connect sensor without delay or replace skin
temperature sensor.
°C HAUT/SKIN/PEAU/PIEL
°C LUFT/AIR
o
°
<
\
'
_J
V
_>
<
_J
7
o
o
o
The intermittent tone can be suppressed for 10 minutes:
Press @ key, yellow
&
LED and
red
Alarm
LED are lit.
27