Component Replacement; Optional Service Mode Setting. P39; Preparation For Component Replacement - BeaconMedaes MAT-S Installation, Operation And Maintenance Instructions

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
6.
Component Replacement
Procedures.
WARNING! It is essential that only genuine
BeaconMedaes
spare
maintenance.
CAUTION! Ensure no contaminates, oil or grease
come into contact with any of the gas connection/
internals.
Become familiarized with all associated procedures and
ensure all required parts are available before carrying
out any of the following procedures.
6.1
Optional Service Mode Setting.
NOTE - This is an optional setting for servicing the
manifold. While in service mode cylinder bank selection
is manual only, and all the alarm outputs are deactivated
and will appear as normal. All manifold indicators will
function as normal.
Figure 32 - Service Mode Option.
1. In the home page
,
push the setting button.
2. In setting page, push
the Advanced button
This step needs password
and only trained service
persons can operate.
3. Click the Setting
of service Mode.
4. Enter the service mode
page.
5. Press the home button
icon
to quit the service mode.
parts
are
fitted
O2
1500
SCFH
1500
56
150
44.0
66.0
PSI
PSI
110.0
.
O2
1500
SCFH
1500
51
150
44.0
66.0
150
PSI
PSI
110.0
SERVICE MODE
or previous
6.2
Preparation for Component
Replacement and Maintenance
Requiring Total Manifold Isolation.
during
6.2.1
Ensure alternative gas supply is available and
functioning before taking the manifold offline.
6.2.2
Remove the front cover by opening the 4 latches
as shown in Figure 33 and lift away.
Figure 33 - Main Cover Removal.
6.2.3
Close the line source isolation valve as shown in
Figure 34 and close all the cylinder valves.
1500
150
PSI
CAUTION! When exhausting anesthetic and
oxygen manifolds ensure that the manifold room is well
ventilated and no potential ignition sources are present.
NOTE - If the cylinder pressure gauges are not rapidly
dropping in pressure, stop draining the system and
check all cylinders are correctly isolated.
Figure 34 - Line Source Isolation Valve.
140
PSI
39
1
Cover
Push
Latch
Button
2
1
2
Pull
to Open
Line
Source
Isolation
Valve in Closed
Position
4109992459.00