Technical Specification. P10 - BeaconMedaes MAT-S Installation, Operation And Maintenance Instructions

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
1.7
Technical Specification.
1.7.1
Technical parameter.
NFPA Technical Parameter
Nominal Delivery / Line Output Pressure (psi)
Nominal Delivery / Line Output Flow (scfh)
2
Stage Regulator Test Pressure as per 10524-2 at 10%
nd
Pressure Drop (psi)
2
Stage Regulator Output Flow (Single) as per 10524-2
nd
at 10% Pressure Drop (scfh)
Input Pressure Range (psi)
1
Stage Regulator Pressure (Intermediate) (psi)
st
2
Stage Regulator Pressure Output (Line) (psi)
nd
High-Pressure Regulator Relief Valve Setting (psi)
Line Pressure Relief Valve Setting (psi)
Changeover Pressure (psi)
1.7.2
Flow performance.
Single, Line 55 psi, Intermediate 127 psi, Inlet 145 psi
70
65
60
55
50
45
40
35
30
25
20
0
93
189.6
Dynamic Pressure (psi)
Single, Line 100 psi, Intermediate 217.5 psi, Inlet 290 psi
120
110
100
90
80
70
60
50
0
337.2
Dynamic Pressure (psi)
4109992459.00
316.2
621
981.6
Line pressure
10% Drop of Sta�c pressure
985.2
1425.6
1489.8
10% Drop of Sta�c pressure
Values
55
2250
57
2000
150-2900
125
55
350
75
145
Single, Line 180 psi, Intermediate 217.5 psi, Inlet 290 psi
220
200
180
160
140
120
100
1050.6
1260.6
0
Dynamic Pressure (psi)
1819.2
Line Pressure
10
100
3200
101
3100
300-2900
220
100
150
290
298.8
370.2
592.8
1423.2
10% Drop of Sta�c pressure
180
3600
183
3300
220
180
250
290
1554.6
1665
1876.2
Line pressure

Advertisement