BeaconMedaes MAT-S Installation, Operation And Maintenance Instructions page 18

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
4.
Press the General icon
5.
Press the value which need to change language type, date
and time. Use the keyboard to modify. Press the save
icon
to save the changes made.
4109992459.00
in the setting page.
Countries
China
English
Season
Summer
Time Mode
24-Hour
Figure 10 - Set Line Pressure.
NOTE - Line pressure will normally be pre-set from
the factory. Options are for nominal 55, 100 & 180 psi
supply. This setting will not affect the actual supply
pressure, it is for setting the default line pressure alarm.
The line pressure setting is changed by adjusting the
regulator knob.
1.
Press the line pressure value and use the keyboard to
change the number. Press the save icon
to save the changes made.
Figure 11 - Set Pressure Units.
NOTE - Changing the pressure units will automatically
convert all pressure related control variable to the newly
selected unit. Options are PSI (default), kPa or bar.
1.
Press the icon
2.
Press the downward arrow
Press the save icon
18
55
1987
44
66
unit of measure.
to change the unit.
to save the changes made.
PSI
SCFH
PSI
PSI
PSI
PSI