Medical Gas Supply System
Safety Precautions
WARNING! DO NOT USE OIL OR GREASE on any
components in Manifold Changeover System for any
reason. This could lead to a FIRE or an EXPLOSION. Only
use approved OXYGEN COMPATIBLE lubricants, which
can be purchased from BeaconMedaes if necessary.
Pressurised gas from the system may cause personnel
injury or property damage if the device is improperly
operated or maintained.
Operator should carefully read and become familiar with
the contents of this manual before install and maintain
the Medical Gas Supply System.
Operator is expected to use common sense safety
precautions, good workmanship practices and follow
any related local safety precautions.
Component descriptions and spare part lists are
available on request.
Identification of symbols
The following symbols apply to this product and are used
in these instructions and on the product in question. The
meanings of these symbols are as specified below: -
Read instructions
Temperature limitation
Humidity limitation
Atmospheric pressure limitation
Date of manufacture
Do not dispose of in general waste
Electromagnetic Interference
The panel has been tested to IEC 61326-1: 2020 Electrical
equipment for measurement, control and laboratory
use –EMC requirements - Part 1: General requirements.
Ensure that all data cables are physically separated from
other mains and data cables.
4109992459.00
Environmental Transport and Storage
Conditions
All products are separately packaged and stored in
controlled conditions.
Minimum environment temperature: -20
Maximum environment temperature: 50
Minimum relative humidity (non-condensing): 10%
Maximum relative humidity (non-condensing): 95%
Environmental Operating Conditions
Adverse environmental conditions and harsh abrasives
or chemicals may cause damage to the unit.
WARNING! Only use approved leak detection
fluids with this product.
Minimum environment temperature: 0
Maximum environment temperature: 40
Minimum relative humidity (non-condensing): 10%
Maximum relative humidity (non-condensing): 95%
Atmospheric pressure range: 700-1100 hpa
Environmental Protection
Discard the unit and/or components in any standard
refuse facility. The unit does not contain any hazardous
substances.
Cleaning
The device should be wiped over with a damp cloth
frequently to remove any dust or foreign substances.
Electrical Details
WARNING! It is necessary to check the integrity
of the power source for safety at regular intervals. These
checks should be carried out annually and replacement
power supplies used as necessary.
Power source - Mains operated using 100-250V, 50/60Hz,
alternating current.
Current requirements - 0.3A
Relative humidity - 95%
Type of protection against electric shock - Class 1 (Mains
supplied equipment using a protected earth).
Mode of operation - Continuous (equipment may be left
switched on indefinitely).