BeaconMedaes MAT-S Installation, Operation And Maintenance Instructions page 20

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
Figure 14 - Alarm Output Option.
1.
Press the Units of Measure
2.
Press the downward arrow
output option. Press the save icon
return to main setting page.
Advanced Set-Up.
1.
Press the icon in the setting page, this action needs
password, only authorized service persons can do that.
2.
Press the of the pressure zeroing to zero the
pressure.
4109992459.00
icon in the settings page.
to change the alarm
PSI
SCFH
O2
3.
Press the
Pressure the to make current pressure follow the
default algorithm in the controller software.
150
4.
Press the of the service mode.
and
In the service mode, one can press the and change-
over button, no matter which alarm sounds.
150
5.
In the advanced page, the gas type and medical standard
also can be selected.
20
to correct current pressure to value 0.
Zero
Default
O2
0
0
44.0
PSI
PSI
O2
1500
1500
51
44.0
PSI
PSI
SERVICE MODE
1500
SCFH
0
66.0
150
PSI
110.0
O2
SCFH
140
66.0
150
PSI
110.0
O2