BeaconMedaes MAT-S Installation, Operation And Maintenance Instructions page 7

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
No. Description
20
Solenoid Valve
21
Smart Box Antenna
22
Manifold Cover
23
Manifold Screen PCB
24
Manifold Power PCB
25
Line Pressure Sensor
26
Line Pressure Relief Exhaust Connection Point
(28 mm OD Copper Tube)
27
Intermediate Pressure Relief Exhaust Connection
Point (12 mm OD Copper Tube)
28
Pipeline Connection Point (22 mm OD Copper
Tube)
Figure 2a - Pneumatic Schematic Diagram.
Master Valve
Non-Return Valve
Header With Non-Return Valve Which
Pigtail
Have Gas Specific Connection
Ball Valve
High Pressure Cylinders
With Isolation Valve
(Left Bank)
Hospital Pipeline
Lockable Ball Valve
Line Pressure
Sensor
Service Valve
Bank
Dome Regulator*
Bleed
Valve
Ball Valve
Ball Valve
Pressure
*
Gauge
Pressure
2nd stage
Gauge
Pressure
Regulator
2nd stage
Pressure
Regulator
Ball Valve
Intermediate
Relief Valve
Bank
Bleed
Non-Return
Valve
Valve
Non-Return
Pressure
Gauge
Sintered
1st stage
Filter
Pressure
Regulator
Pressure
Pressure
Gauge
Silencer
Sensor
MAT-S MANIFOLD
* NOTE: REQUIRED WHEN DELIVERY PRESSURE IS 100 PSI TO 180 PSI
Symbols to BS2971 & ISO 1219-1
Relief Valve Exhaust
Vent Valve
Line Relief Valve
Bank
Bleed
Valve
Pressure
Gauge
Ball Valve
Intermediate Relief Valve
Bank
Reserve
Bleed
Source
Valve
Of Supply
Valve
Pressure Gauge
1st stage
Sintered
Pressure
Filter
Regulator
Solenoid Valve
Pressure
Pressure
(Bank Selection)
Sensor
Gauge
Left Bank Is Default
Working Bank
High Pressure Cylinders
With Isolation Valve
(Right Bank)
7
Master Valve
Non-Return Valve
Header With Non-Return Valve Which
Pigtail
Have Gas Specific Connection
Ball Valve
Main Gas Line
Control Gas Line
Exhaust Line