Medical Gas Supply System
6.

Component Replacement

Procedures.
WARNING! It is essential that only genuine
BeaconMedaes
spare
maintenance.
CAUTION! Ensure no contaminates, oil or grease
come into contact with any of the gas connection/
internals.
Become familiarized with all associated procedures and
ensure all required parts are available before carrying
out any of the following procedures.
6.1
Optional Service Mode Setting.
NOTE - This is an optional setting for servicing the
manifold. While in service mode cylinder bank selection
is manual only, and all the alarm outputs are deactivated
and will appear as normal. All manifold indicators will
function as normal.
Figure 35 - Service Mode Option.
1. In the home page
,
push the setting button.
2. In setting page, push the
Advanced button
.
3. Click the Setting
of service Mode.
4. Enter the service mode
page.
5. Press the home button
icon
to quit the service mode.
parts
are
fitted
SERVICE MODE
or previous
6.2
Preparation for Component
Replacement and Maintenance
Requiring Total Manifold Isolation.
during
6.2.1
Ensure alternative gas supply is available and
functioning before taking the manifold offline.
6.2.2
Remove the front cover by opening the 4 latches
as shown in figure 36 and lift away.
Figure 36 - Main Cover Removal.
6.2.3
Close the line isolation valve as shown in figure
37 and close all the cylinder valves.
6.2.4
Open the test point isolation valve as shown
in figure 37 and probe the outlet to depressurise the
system.
CAUTION! When exhausting anesthetic and
oxygen manifolds ensure that the manifold room is well
ventilated and no potential ignition sources are present.
NOTE - If the cylinder pressure gauges are not rapidly
dropping in pressure, stop draining the system and
check all cylinders are correctly isolated.
Figure 37 - Valve Isolation and Pressure Draining.
33
1
Cover
Push
Latch
Button
2
1
2
Pull
to Open
Line
Isolation
Valve in Closed
Position
Test Point
Isolation Valve in
Open Position
Sampling
Outlet, Gas
Venting Point
4109992485.00