BeaconMedaes MAT-S Installation, Operation And Maintenance Instructions page 19

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
Figure 12 - Cylinder Set-Up.
NOTE - Minimum requirement in the cylinder set-up is
to set the correct maximum pressure. This will correctly
scale the graphic on the display. The size and number
of cylinders per bank is for calculating the average gas
usage only.
1.
Press the system icon
2.
Press the Cylinder Pressure and Cylinder Volume to modify
the value by the keyboard. Press the save icon
to save the changes made.
3.
Press the NO. Cyl (Left) and NO. Cyl (Right) to modify the
quantity by keyboard. Press the save icon
to save the changes made.
NOTE - The typical average gas usage is only for gas
phase in the cylinders. This calculated value will not
convert correctly for liquid phase in the cylinders.
in the setting page.
55
PSI
1987
PSI
44
PSI
PSI
66
55
PSI
1987
PSI
44
PSI
PSI
66
Figure 13 - Line Pressure Alarm Set-up.
NOTE - Default Line Pressure Alarm setting will be
automatically defined when the Nominal Line Pressure
is selected. See Table 3 below for default values. These
values can be manually adjusted as required to suit on
site requirements.
1.
Press the system icon
2.
Press the Low Line Pressure Alarm and High Line Pressure
Alarm and modify the value by keyboard. Press the save
icon and return to main setting page.
Table 3: Default Line Pressure Alarm Settings and
Adjustment Range.
Nominal Line
Low Pressure
Pressure
55 psi
100 psi
180 psi
NOTE - Low line pressure set-point may need to be reduced
if the device is used as a backup manifold. See Table 5 in
Section 4 for reference.
19
in the settings page.
55
PSI
1987
PSI
44
PSI
66
PSI
Default Values
High Pressure
Alarm
Alarm
44
66
80
120
144
216