• To prevent the uncontrolled flow of medication, use a CADD
Anti-Siphon Valve, a CADD
®
or a CADD
Administration Set with an attached Add On Anti-Siphon Valve.
• Use of a syringe with the CADD
DELIVERY of medication. Syringe function can be adversely affected by varia-
tions in plunger dimension and lubricity, which can result in greater force being
required to move the plunger. A syringe will lose plunger lubrication as it ages
and, as a result, the amount of under-delivery will increase which could, on occa-
sion, be significant. Therefore, the type of medication therapy and delivery accu-
racy required must be considered when using a syringe with the CADD
Clinicians must regularly compare the volume remaining in the syringe to the
pump's displayed values such as RES VOL and GIVEN in order to determine
whether under-delivery of medication is occurring and, if necessary, take appro-
priate action.
2.1.2

CAUTIONS

• This device is not intended to be used for delivery of blood or cellular blood
products.
• This device may interfere with ECG equipment. Monitor ECG equipment care-
fully when using this device.
• The pump is not sterile. It is not designed to be sterilized. Sterilization could
damage the microcomputer and other pump parts.
• The pump should be routinely cleaned and kept free of dirt, liquids, and foreign
objects.
• Do not store the pump at temperatures below -40°C (-40°F ) or above 55°C
(131°F).
• Do not operate the pump at temperatures below +2°C (35°F) or above 40°C
(104°F).
• Do not expose the pump to humidity levels above 90% R. H.
• The pump is water resistant. However, total immersion is not recommended be-
cause moisture buildup within the case may damage the parts. Do not use pump
in the shower, sauna, or steam bath.
• Do not store the pump for prolonged periods with a battery; the battery could
leak and damage the pump.
• Avoid using the pump in close proximity to sources of strong static electricity or
strong electromagnetic fields.
• The use of a Deltec Pump Pouch is recommended. If the pump is dropped or
inadvertently hit against a hard surface, the pump pouch is designed to minimize
the need for servicing.
®
Administration Set with integrated anti-siphon valve,
®
Administration Set may result in UNDER-
Extension Set with
®
®
pump.