will not occur, and the pump will begin its delivery immediately. For example,
if it is 4:30 pm, and you do not want the delivery of the drug to begin until 6:00
pm, you would program the DELAY START for 1HR:30MIN, which means
that the delivery would not begin until 6:00 pm.
2.5.7

The CYCLE GIVEN (CYCLE GIVEN)

CYCLE GIVEN shows the time that has elapsed since the last dose began.
CYCLE GIVEN is automatically set to 00:00 when the clinician programs a
new drug delivery profile or a new DELAY START time. When the pump is
started, CYCLE GIVEN waits until the DELAY START time, if any, passes.
Then it begins to count in 1-minute increments (or 1-hour increments in the 24-
hour mode) until it equals the programmed delivery CYCLE. At that point,
CYCLE GIVEN automatically resets to 00:00 and a new cycle begins.
If the pump is stopped at any time during the cycle, CYCLE GIVEN stops
counting. When the pump is restarted, CYCLE GIVEN resumes counting at the
same point in the cycle at which it was stopped.
NOTE:
If necessary, you can reset the CY CL E G IV EN in L L 0 or L L 1. To do this, display
the D EL AY STA RT screen and press SET /CL EA R . Even if the D EL AY STA RT
is 00:00, the CY CL E G IV EN w ill clear to 00:00 and a new cycle w ill begin w hen
the pum p is restarted.
2.5.8

The Amount GIVEN (GIVEN)

GIVEN refers to the number of milliliters that have been delivered since the
GIVEN register was last cleared.
The number of milliliters in GIVEN accumulates from 000 to 999 and then
returns to 000. Please note that the GIVEN counter functions in a way that is
similar to an automobile's odometer. After the GIVEN counter reaches 999, it
will start at 000 again and continue counting, which is especially important if
you plan to use more than 1,000 milliliters of solution.
11