INSTRUCTION MANUAL
INJECTION UNIT
For LIGHTNING SURGE SIMULATOR
LSS-6330 / LSS-F03 SERIES
MODEL: LSS-INJ6401SIG
NOISE LABORATORY CO., LTD.
1.02 Edition
AEE00585-00E-0

Advertisement

Table of Contents
loading

Summary of Contents for NoiseKen LSS-6330 Series

  • Page 1 INSTRUCTION MANUAL INJECTION UNIT For LIGHTNING SURGE SIMULATOR LSS-6330 / LSS-F03 SERIES MODEL: LSS-INJ6401SIG NOISE LABORATORY CO., LTD. 1.02 Edition AEE00585-00E-0...
  • Page 2 NOTICE The contents of this booklet are subject to change without prior notice. No part of this booklet may be reproduced or transferred, in any form, for any purpose, without the permission of Noise Laboratory Co., Ltd. The contents of this booklet have been thoroughly checked. However, if a doubtful point, an error in writing or a missing is found, please contact us.
  • Page 4: Important Safety Precautions

    5. When connecting this unit to accessories for test waveforms and test conditions, optional equipment and other equipment, press STOP button of the lightening surge simulator (LSS-6330 series) and check that lamp of STOP button is lighting up beforehand. Otherwise, you may receive an electric shock. There may be residual voltage even if the test is stopped.
  • Page 5: Accessories

    2. ACCESSORIES A B C D E F G H I A: Accessory bag ..................1 pc B: Surge In Cable (1m) ................2 pcs C: Line Out Cable (1m) ................5 pcs D: Connection Plug ................... 7 pcs E: Capacitor / Gas Arrestor Unit .............. 1 pc F: Gas Arrestor Unit ..................
  • Page 6: Application Form For Instruction Manual

    3. APPLICATION FORM FOR INSTRUCTION MANUAL We place an order for an instruction manual. Model: LSS-INJ6401SIG Serial No.: Applicant: Company name: Address: Department: Person in charge: Tel No.: Fax No. Cut off this page "APPLICATION FORM FOR INSTRUCTION MANUAL" from this volume and keep it for future use with care. When an INSTRUCTION MANUAL is required, fill in the above Application Form and mail or fax it to the following sales department of our company.
  • Page 7 Memorandum...
  • Page 8: Table Of Contents

    4. TABLE OF CONTENTS 1. IMPORTANT SAFETY PRECAUTIONS ........................1 2. ACCESSORIES ................................2 3. APPLICATION FORM FOR INSTRUCTION MANUAL ..................3 4. TABLE OF CONTENTS ............................... 5 5. PREFACE ..................................6 ..................................... 6 5-1. P REFACE ..................................... 6 5-2. F EATURE 6.
  • Page 9: Preface

    5. PREFACE 5-1. Preface We thank you very much for your purchase of our injection unit for the Lightening Surge Simulator for LSS-6330 / LSS-F03 SERIES (LSS-INJ6401SIG). The unit is the injection unit for tests for interconnection lines described in IEC61000-4-5. Please fully read and utilize this manual before using this unit.
  • Page 10: Basic Safety Precautions

    As the unit outputs high voltage (6600V maximum) and great current (157A maximum) by surge from LSS-6330 series, carefully handle it. Mishandling or careless operation may result in a fatal wound. Do not touch the surge in connector, DC line out connector, DC line in terminal in conducting tests because high voltage is generated.
  • Page 11 14. Do not open the unit because high voltage is generated internally. 15. In case of bypassing the inductor of each line’s decoupling network, as the surge can be back to the input side with some EUT impedance condition, connecting protection instruments is recommendable.
  • Page 12: Appearance Of The Unit

    7. APPEARANCE OF THE UNIT FRONT PANEL INTERCONECTION LINE UNIT LSS-INJ6400SIG SURGE RETURN SURGE INJ LINE IN LINE SELECT LINE SELECT OUT1 SURGE PROTECTION OUT2 OUT3 OUT4 INDUCTORS BY-PASS CAPACITORS ARRESTORS COUPLING COUPLING SURGE IN HOT LINE OUT1 LINE OUT2 LINE OUT3 LINE OUT4 LINE OUT GND SURGE IN COM SURGE IN 6.6kV MAX...
  • Page 13 REAR PANEL INTERLOCK L I N E I N 1 2 3 4 1 2 3 4 G ⑧DC LINE IN TERMINAL [LINE IN]: Inputs power for EUT. ⑨INTERLOCK CONNECTOR [INTERLOCK: 1,2]: Connect [1] to the lightening surge simulator’s interlock connector by the interlock connection cable (accessory). Connect interlock connector (of the surge simulator’s accessory) to [2].
  • Page 14: How To Connect Equipment

    8. HOW TO CONNECT EQUIPMENT 8-1 Connection of Rear Panel Connect PE terminal to FG terminal of LSS series by PE cable (accessory) as indicating in the figure below. Connect the interlock connector [1] to the LSS series interlock connector by the interlock cable (accessory).
  • Page 15: Connection Of The Front Panel (In Case Of Lss-F03 Series)

    8-3. Connection of the front panel (in case of LSS-F03 series) Connection of the LSS-INJ6401SIG to the LSS-F03 is done by using the Surge In Cable(05-T1578) sold separately. Connect the safety socket side of the cables to the SURGE IN terminals (HOT,COM) on the LSS-INJ6401SIG and High Voltage terminal side to the SURGE OUTPUT (HOT,COM) terminals on the LSS-F03.
  • Page 16: Setting When Arrestors Coupling

    8-4 Setting When Arrestors Coupling (1) To inject surge to Line - Line Example: In case of injection line: [OUT1] and return line: [OUT2] SURGE RETURN SURGE INJ LINE IN LINE SELECT LINE SELECT OUT1 SURGE PROTECTION OUT2 OUT3 OUT4 INDUCTORS BY-PASS CAPACITORS...
  • Page 17 (2) To inject surge to Line-Ground Example: In case of injection line: [OUT1] and return line: [GND]. SURGE RETURN SURGE INJ LINE IN LINE SELECT LINE SELECT OUT1 SURGE PROTECTION OUT2 OUT3 OUT4 INDUCTORS BY-PASS CAPACITORS ARRESTORS COUPLING COUPLING SURGE IN HOT LINE OUT1 LINE OUT2 LINE OUT3 LINE OUT4 LINE OUT GND SURGE IN COM SURGE IN...
  • Page 18: Setting When Capacitors Coupling

    8-5 Setting When Capacitors Coupling (1) To inject surge to Line - Line Example: In case of injection line: [OUT1] and return line: [OUT2] SURGE RETURN SURGE INJ LINE IN LINE SELECT LINE SELECT OUT1 SURGE PROTECTION OUT2 OUT3 OUT4 INDUCTORS BY-PASS CAPACITORS...
  • Page 19 (2) To inject surge to Line-Ground Example: In case of injection line: [OUT1] and return line: [GND]. SURGE RETURN SURGE INJ LINE IN LINE SELECT LINE SELECT OUT1 SURGE PROTECTION OUT2 OUT3 OUT4 INDUCTORS BY-PASS CAPACITORS ARRESTORS COUPLING COUPLING SURGE IN HOT LINE OUT1 LINE OUT2 LINE OUT3 LINE OUT4 LINE OUT GND SURGE IN COM SURGE IN...
  • Page 20: Block Chart

    9. BLOCK CHART アクリル扉部インターロックスイッチ INTERLOCK of the door INTER LOCK1 SURGE IN 40Ω CAPACITORS ARRESTORS INTER LOCK2 COUPLING COUPLING 0.5μF SURGE INJ. LINE SELECT OUT4 OUT3 OUT2 OUT1 SURGE RETURN LINE SELECT LINE IN (DC50V/1A) LINE OUT INDUCTORS BY-PASS SURGE SURGE PROTECTION PROTECTION...
  • Page 21: Specifications

    10. SPECIFICATIONS Surge Generating Part ITEM SPECIFICATIONS REMARKS Maximum coupling Up to 6.6kV surge voltage Maximum coupling Up to 144A(capacitor coupling) surge current Up to 157A(arrestor coupling) Injection mode line-line/line-PE Surge waveform specification(capacitor coupling) ITEM SPECIFICATIONS REMARKS Surge waveform Open voltage 4kV ±10% Please use high voltage cable 05-T1578 (sold...
  • Page 22: Warranty

    11. WARRANTY Servicing terms The following terms are applicable to servicing by Noise Laboratory Co., Ltd., (hereafter referred to as the Company) provided to maintain the intended performance of its products. 1. Scope The following terms shall apply only to products made by the Company. 2.
  • Page 23 Limited warranty Noise Laboratory Co., Ltd. (hereafter referred to as the Company) warrants its products to be free from defects in materials and workmanship under normal use and service for a period of one year from date of delivery. In the event of failure of a product covered by this warranty, the Company will repair the product or may, at its option, replace it in lieu of repair without charge.
  • Page 24: Maintenance

    12. MAINTENANCE 1. When repair, maintenance or internal adjustment of the unit is required, a qualified service engineer takes charge of such work. 2. Maintenance on the user side is restricted to the outside cleaning and functional check of the unit. 3.
  • Page 25: Noise Laboratory Support Network

    13. NOISE LABORATORY SUPPORT NETWORK ⚫ If a symptom which seems a trouble is found, inform the model name and serial number of the product together with the symptom to Noise Laboratory or your nearest sales agent of Noise Laboratory. ⚫...
  • Page 27 NOISE LABORATORY CO., LTD. 1-4-4, Chiyoda, Chuo-ku, Sagamihara City, Kanagawa Pref., 252-0237, Japan TEL: +81-(0)42-712-2051 FAX: +81-(0)42-712-2050 URL: http://www.noiseken.co.jp Printed in Japan...