Emc Information - Eizo RadiForce MS235WT Instructions For Use Manual

Touch panel color lcd monitor
Table of Contents

Advertisement

EMC Information

Essential performance of RadiForce series is to display images and operate functions normally.
CAUTION
The RadiForce series requires special precautions regarding EMC and need to be installed, put into
service and used according to the following information.
Do not use any cables other than the cables that provided or specifi ed by us.
Using other cables may cause the increase of emission or decrease of immunity.
Do not put any portable and mobile RF communications equipment close to the RadiForce series.
Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation
in the confi guration in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts,
confi guring a medical system, is responsible that the system complies with the requirements of IEC/
EN60601-1-2.
Guidance and manufacturer's declaration - electromagnetic emissions
TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspecifi  edbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.
Emission test
Compliance Electromagnetic environment - guidance
RFemissions
Group1
CISPR11/EN55011
RFemissions
ClassB
CISPR11/EN55011
Harmonicemissions
ClassD
IEC/EN61000-3-2
Voltagefl  uctuations/
Complies
fl  ickeremissions
IEC/EN61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspecifi  edbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.
Immunity test
IEC/EN60601 test
level
Electrostatic
±6kVcontact
discharge(ESD)
±8kVair
IEC/EN61000-4-2
Electricalfast
±2kV
transient/burst
forpowersupplylines
IEC/EN61000-4-4
±1kV
forinput/outputlines
Surge
±1kVline(s)toline(s)
IEC/EN61000-4-5
±2kVline(s)toearth
Voltagedips,short
<5%U
interruptionsand
U
)for0.5cycle
T
voltagevariationson
40%U
powersupplyinput
for5cycles
lines
70%U
IEC/EN61000-4-11
for25cycles
<5%U
U
)for5sec
T
Powerfrequency
3A/m
(50/60Hz)
magneticfi  eld
IEC/EN61000-4-8
NOTEU
isthea.c.mainsvoltagepriortoapplicationofthetestlevel.
T
58
Appendix
TheRadiForceseriesusesRFenergyonlyforitsinternalfunction.
Therefore,itsRFemissionareverylowandarenotlikelytocauseany
interferenceinnearbyelectronicequipment.
TheRadiForceseriesissuitableforuseinallestablishments,includingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Compliance level
±6kVcontact
±8kVair
±2kV
forpowersupplylines
±1kV
forinput/outputlines
±1kVline(s)toline(s)
±2kVline(s)toearth
(>95%dipin
<5%U
(>95%dipin
T
T
U
)for0.5cycle
T
(60%dipinU
)
40%U
(60%dipin
T
T
T
U
)for5cycles
T
(30%dipinU
)
70%U
(30%dipin
T
T
T
U
)for25cycles
T
(>95%dipin
<5%U
(>95%dipin
T
T
U
)for5sec
T
3A/m
Electromagnetic environment - guidance
Floorsshouldbewood,concreteorceramictile.
Iffl  oorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheRadiForceseriesrequirescontinued
operationduringpowermainsinterruptions,it
isrecommendedthattheRadiForceseriesbe
poweredfromanuninterruptiblepowersupplyor
abattery.
Powerfrequencymagneticfi  eldsshouldbeat
levelscharacteristicofatypicallocationina
typicalcommercialorhospitalenvironment.