1.2 MFC 300F & MFC 300F T6
The MFC 300F & MFC 300F T6 have intrinsically safe connections to the mass flow sensor with either
increased safety or intrinsically safe signal outputs. The signal output connection compartment can be
configured with protection type Ex d or Ex e. The marking is as follows:
For Ex i outputs
Ex d connection compartment
II 2(1) G Ex d [ia/ib] IIC T6 Gb
II 2(1) D Ex ib t [ia Da] IIIC T80°C Db
For non-Ex i outputs
II 2 G Ex d [ib] IIC T6 Gb
II 2 D Ex ib t IIIC T80°C Db
The output connections to the mass flow sensor have the following values:
MFC 300F
Uo = 16.5 V
Io = 305 mA
Po = 1.25 W
Co = 230 nF
Lo = 320 μH
For details of the signal output values, compare the table in section 4 with the CG32.......XYZ
number listed on the data label.
6
Ex e connection compartment
II 2(1) G Ex de [ia/ib] IIC T6 Gb
II 2(1) D Ex ib t [ia Da] IIIC T80°C Db
II 2 G Ex de [ib] IIC T6 Gb
II 2 D Ex ib t IIIC T80°C Db
Power Supply Circuit
MFC 300F T6
Uo = 16.5 V
Io = 262 mA
Po = 1.07 W
Characteristic curve: linear
Co = 240 nF
Lo = 450 μH
Data Circuit
MFC 300F / MFC 300F T6
Uo = 6 V
Io = 33 mA
Po = 120 mW
Co = 1.9 μF
Lo = 2 mH