Measuring Pressure - Omega PCL1200 User Manual

Portable, high accuracy multifunction calibrator
Table of Contents

Advertisement

3.3-2 Using Resistance-Temperature-Detectors (RTDs)
The supported types of RTDs are shown in Section 8.
Specifications. RTDs are characterized by their 0°C resistance,
R0. The calibrator accepts two, three, and four wire inputs, with
four wire input being the most accurate.
To use the RTD option, apply the following steps:
1.
2.
3.
4.
5.
Figure 8. Measuring Temperature with RTD Connection
Resistance can also be measured using this function. To do so,
use the above procedure and choose OHMS from the sensor
types. This option can be used along with a table to measure an
RTD which is not programmed into the calibrator.

3.4 Measuring Pressure

Note: The Omega Pressure Module connector 700mA needs to be
purchased to connect pressure module to calibrator.
14
Switch to lower display [LOWER] from Main Menu.
Select RTD from the primary parameters. Select [IN] from
input/output control.
Choose 2, 3, or 4-wire connection [2W, 3W, 4W]. (4-wire
allows for the most precise measurement)
Select RTD type from the sensor types.
Attach RTD leads as shown in Figure 8.