Freq_Unit; Func; Local; Lower_Meas - Omega PCL1200 User Manual

Portable, high accuracy multifunction calibrator
Table of Contents

Advertisement

FREQ_UNIT

Sets the unit for frequency. There are three ranges of frequencies for
frequency and pulse modes, CPM (cycles per minute), Hz, and kHz.
Use this command to select the right range. For example:
FREQ_UNIT HZ sets the frequency to Hz range
FREQ_UNIT?
Returns the frequency unit currently being used by the frequency
and pulse modes.

FUNC?

Returns the current mode of the upper and lower displays. For
example if the calibrator is set to volts on the upper display, and
pressure on the lower display, FUNC? Would return:
DCV, PRESSURE

LOCAL

Restores the calibrator to local operation if it was in remote mode.
Also clears LOCKOUT if the unit was in lockout mode.

LOWER_MEAS

Sets the lower display to measure mode. The command is followed
by any of the parameters except for pulse and mA sim, which are
source only. Enter DCI for mA, DCV for volts, TC for thermocouple,
RTD for RTD, FREQUENCY for frequency, and PRESSURE for
pressure. For example:
LOWER_MEAS DCV sets the lower display mode to VOLTS IN

L_PRES_UNIT

Sets the unit for measuring pressure on the lower display. Add the
unit after the command. The available pressure units and their syntax
are shown in the Table 7. (Parameter Units).
For example:
L_PRES_UNIT KPAL sets the pressure unit to kiloPascals
45

Advertisement

Table of Contents
loading

Table of Contents