Operation
5.4 Protective conductor resistance measurement
Procedure for continuous measurement
1. Connect the 4 mm safety measuring lines to the device [} page 25].
2. Only for new safety measuring lines: Please carry out a null balance [} page 27].
3. Bring the 4 mm safety measuring lines into contact with the measuring points.
4. Press and hold the "R
5. The measured protective conductor resistance R
display.
6. To stop the continuous measurement, press the "R
5290 / 06/2023 en
" key until the "R
 LOCK" symbol is displayed.
PE
PE
BENNING PV 1-1+
is shown continuously on the digital
PE
" key.
PE