The product can be used in such measurements: HI and LO input terminals connect to mains in
these devices (maximum 300VAC) or to the branch outlets. However, the HI and LO input
terminals of this product can not be connected to mains in permanently-installed electric devices,
such as main circuit-breaker panel, sub-panel disconnect boxes, or permanently wired motors.
Such devices and circuits are easily subject to overvoltages that may exceed the protection limits
of the product.
Note: Voltages above 300 VAC may be measured only in circuits that are isolated from mains.
However, transient overvoltages are also present on circuits that are broken from mains. The
product is designed to safely withstand occasional transient overvoltages up to 2500 Vpk. Do not
use this equipment to measure circuits where transient overvoltages could exceed this level.
Environmental Precautions
The product complies with requirements indicated in WEEE Directive (2002/96/EC).The affixed
product label (see below) indicates that you must not discard this electrical/electronic product
along with domestic household waste.
Product category: With reference to the equipment types specified in the WEEE directive
Annex 1, this product is classified as "Monitoring and Control Instrument" product.
Some substances contained in the product may be harmful to environment or human body. In
order to avoid hazardous substances releasing to environment or endangering human body, it is
recommended to recycle the product in a proper way to make sure that most of materials can be
correctly re-used or recycled. Contact local authority for disposal or recycling information.
Symbols on Product
The following symbols may appear on products: