Resistance Measurement - East Tester ET3260 User Manual

6 1/2 digit multimeter
Table of Contents

Advertisement

Users manual

2.1.6 Resistance measurement

Basic information
10Ω, 100Ω, 1kΩ, 10kΩ, 100kΩ, 1MΩ, 10MΩ, 100MΩ and 1GΩ.
Maximum resolution: 10uΩ.
Operation description
1) Press
to select 2-wire resistance measurement or 4-wire resistance measurement
function.
Description: Major displayed value refers to measurement voltage value, and auxiliary displayed
value refers to the ratio between the current measurement value and full measurement range.
2) Connect test lines as indicated in the chapter of 1.4 Measurement wiring.
3) Select measurement range as required, and may select automatic measurement, or press the
softkeys of Measurement range+ (
measurement range, which have same effect as keys of
4) Press the softkey of Relative (
difference between actual measurement value and current recorded value.
5) Press the softkey of Display (
interface for details.
6) If necessary, press
and zero calibration and refer to the chapter of Measurement configuration for details. Start
automatic zero calibration during 4-wire resistance measurement.
ET3260 and ET3260A
) and Measurement range- (
) to record the current value, after which display the
) to switch modes, and refer to the chapter of User
to configure the following measurement parameters: integral time
6 1/2 Digit Multimeter
and
.
20
) to select fixed

Advertisement

Table of Contents
loading

This manual is also suitable for:

Et3260a

Table of Contents