Resolution - East Tester ET3260 User Manual

6 1/2 digit multimeter
Table of Contents

Advertisement

Users manual
Input resistance setting is stored in volatile memory, and after cutting off power or resetting
remote interface, multimeter will reelect 10Ω (including all the DC voltage shifts).
Front panel operation: After starting DC voltage measurement function, press
the softkey of Impedance (
key. Default value is 10MΩ.

3.1.4 Resolution

Resolution is expressed by digits which can be measured or displayed by multimeter, which can
be set as 4, 5 or 6 integrated digits, and add [1/2] single units represented by [0] or [1]. If it is
expected to increase measurement accuracy or control noise, please select 6 1/2 digits, while to
increase measurement speed, please select 4 1/2 digits.
Resolution setting is suitable for all the measurement functions. Resolution of math operation
(zero-digit value, limits, dB, dBm and limit test) is the same as that of measurement function.
Corresponding relation between resolution digit and integral time (within power period) is as
follows:
Resolution
High speed 4 1/2
Low speed 4 1/2
High speed 5 1/2
Low speed 5 1/2
High speed 6 1/2
Mid speed 6 1/2
Low speed 6 1/2
Resolution is stored in volatile memory, and after cutting off power or resetting remote
interface, multimeter will set resolutions of all the measurement functions as mid speed 6
1/2.
During DC continuity and diode test, resolution is fixed as high speed 5 1/2.
During DC and resistance measurement, change of resolution digit number does not only
change multimeter resolution, but also integral time, namely, during one certain
measurement, input signal period in sampling of simulation/digit converter (A/D Converter)
on multimeter. Meanwhile, please refer to the chapter of [Integral Time].
During AC measurement, actually resolutions are all fixed as 6 1/2 digits.
During ratio measurement, resolution refers to the resolution of receiving signals at input
terminal.
During front panel operation, there are two methods to change resolution.
1. During measurement function, keys of
2. After pressing
, press the softkey of Integral (
keys of
and
, and then press
ET3260 and ET3260A
), and
and
and
.
6 1/2 Digit Multimeter
to input resistance, and then press
Integral Time
0.02 PLC
0.1PLC
0.2PLC
1PLC
2PLC
10PLC
100PLC
can change resolution.
), select different integral times through