Trigger Parameter Setting; Trigger Output - East Tester ET3260 User Manual

6 1/2 digit multimeter
Table of Contents

Advertisement

Users manual
The sensitivity range is one of the following values, including 0.01% and 0.1% (default value),
1.00% or 10.00% of the reading. For example, if you select 1.00% range, input signal of
multimeter is 5.000V, and only if three continuous readings are all from 4.975 to 5.025, can
multimeter capture and display new readings.
What calls for special attention is that if the new reading is within the sensitivity range of the
current displayed value, multimeter will not update displayed value even though it is determined
as stable reading. For example, the current displayed value is 5.000V, and the next continuous
three measurements are all 5.498V, the screen still displays 5.000V.
In measurement status, press
lights on, and press
If it is expected to modify sensitivity range, press
interface to select needed sensitivity.

3.3.3 Trigger parameter setting

Press
(Setup) to enter trigger parameter setting interface to set samples and
trigger delay.
Samples:
Every time multimeter receives trigger signal, it will read readings in specified number.
The selected samples are stored in volatile memory, and multimeter will set the samples as 1
after cutting off power or resetting remote interface.
Press the softkey of Number (
default value is 1.
Delay:
You can add delay time to trigger signal and each sample behind it, which is quite useful to settle
down input signal or adjust interval between a series of readings and other applications before
you want to capture readings.
Trigger delay time is stored in volatile memory, and multimeter will select auto trigger
delay after cutting off power or resetting remote interface. Default value is 0.
If you select non-auto setting status, the delay time will be suitable for all the measurement
functions and measurement ranges.
If multimeter is configured to capture more than one readings in every trigger, the trigger
delay time specified by you will be inserted between trigger signal and each reading.
Press the softkey of Delay (
values and unit through direction keys. Setting range is from 1 to 3600s, and default value is 0.

3.3.4 Trigger output

Press
(Output) to enter configuration interface of trigger output function, and you
can select to start or stop.
ET3260 and ET3260A
to start holding function, Holding indicator in status column
to stop holding function.
) to set samples, wherein setting range is from 1 to 50000 and
) to enter delay configuration setting interface, and modify
6 1/2 Digit Multimeter
34
(Remain) to enter setting