Technical specifications
7.4 Technical specifications CALOMAT 6F
Parts in gas path wetted by sample gas
Tubing
Sample chamber
Gas path made of pip‐
Gas bushing
ing
Piping
Sample chamber
Special applications
Gas path made of pip‐
Gas bushing
ing
Pipe/nozzle
Sample chamber
Gaskets (O-rings)
Further designs upon request
Gas inlet conditions
Permissible sample gas pressure
Purging gas pressure
Sample gas flow
Sample gas temperature
•
Unheated devices
•
Heated devices
Sample gas humidity
7.4
Technical specifications CALOMAT 6F
Design, enclosure
Weight (analyzer)
Electrical characteristics
Power consumption
Fuse values
•
100 to 120 V
•
200 to 240 V
Gas path
Gas connections
Gaskets (O-rings)
Sensor
60
FKM
Stainless steel 1.4571 or tantalum
Titanium or stainless steel 1.4571
or Hastelloy C22
Titanium or stainless steel 1.4571
or Hastelloy C22
Stainless steel 1.4571 or tantalum
Hastelloy C22
Hastelloy C22
Hastelloy C22
Material according to specification
500 to 3000 hPa (7.3 to 43.5 psi) absolute
< 165 hPa (2.4 psi) over ambient pressure
18 to 60 l/h (0.3 to 1 l/min)
0 to 50 °C (32 to 122 °F)
Max. 15 °C (27 °F) above temperature of analyzer part
< 90% RH (relative humidity), non-condensing
Approx. 25 kg (55 lbs.)
Approx. 30 VA
1T/250
0.63T/250
Stainless steel 1.4571, pipe diameter 1/4" or 6 mm
FFKM (e.g. Kalrez)
Silicon, SiO
N
, gold, epoxy resin, glass
x
y
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
Compact Operating Instructions, 09/2018, A5E45779389002-01
--
--
Titanium or stainless steel 1.4571
or Hastelloy C22
Titanium or stainless steel 1.4571
or Hastelloy C22
Stainless steel 1.4571 or tantalum
Hastelloy C22
Hastelloy C22
Hastelloy C22
Material according to specification