Specifications - Horiba Scientific LAQUAtwin B-771 Instruction Manual

Compact conductivity meter
Table of Contents

Advertisement

8.2

Specifications

Model
Measurement principle
Minimum sample volume
Measurement
range
Display range
Range and
resolution (valid digits)
Calibration
Accuracy
Display
Operating temperature/
humidity
Power
Battery life
Main materials
Outer dimensions/
mass
Main functions
*1
When the measured value is out of the measurement range, the displayed
value blinks. It should be used only as a guide.
*2
Selectable between one-point and two-point calibrations.
The high conductivity standard solution (12.9 mS/cm) is sold separately.
*3
Repeatability in measurement of a standard solution after calibration using the
same standard solution.
(1) ±5 S/cm (0 S/cm to 199 S/cm), (2) ±0.05 mS/cm (0.20 mS/cm to 1.99
mS/cm), (3) ±0.5 mS/cm (2.0 mS/cm to 19.9 mS/cm), (4) ±5 mS/cm (20 mS/
cm to 199 mS/cm)
*4
IP67: no failure when immersed in water at a depth of 1 meter for 30 minutes.
But the product cannot be used underwater.
B-771
2 AC bipolar method
0.12 mL or more
Conductivity
0 mS/cm to 19.9 mS/cm, 0 S/m to 1.99 S/m
Salt
0% to 1.1%
TDS
0 ppm to 9900 ppm
0 mS/cm to 199 mS/cm (0 S/m to 19.9 S/m)
(1) 0 S/cm to 199 S/cm: 1 S/cm
(2) 0.20 mS/cm to 1.99 mS/cm: 0.01 mS/cm
(3) 2.0 mS/cm to 19.9 mS/cm: 0.1 mS/cm
(4) 20 mS/cm to 199 mS/cm: 1 mS/cm
Two-point
±2% full scale ±1 digit (for each range)
Custom (monochrome) digital LCD
5C to 40C,
85% or less in relative humidity (no condensation)
CR2032 batteries (2)
Approx. 400 hours in continuous use
ABS epoxy
164 mm × 29 mm × 20 mm (excluding projections)
Approx. 45 g (without batteries)
Temperature conversion (2%/C fixed), waterproof
reading locking, automatic power OFF
*2
22
*1
*3
*4
,