One-Point Calibration - Horiba Scientific LAQUAtwin B-771 Instruction Manual

Compact conductivity meter
Table of Contents

Advertisement

5.2

One-point calibration

1.
Set "One-point calibration" (refer to page 18).
2.
Open the protection cover and put some drops
of the 1.41 mS/cm standard solution into the
measurement cell.
Washing the sensor with the standard solution
beforehand may provide more accurate calibration.
3.
Close the protection cover and press the CAL
switch for over 2 seconds.
and
displayed.
After the calibration is completed,
stop blinking and light steadily.
4.
Clean the sensor with tap water and
remove moisture.
5.
Press the MEAS switch for 0.5 seconds to
enter the measurement mode and prepare
for measurement.
blink and the calibration value is
and
10