Instructions for use EXPERTmatic LUX E25 L Planmeca – 1.015.0660
6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting
4 Press the O-ring between your fingers to form a loop.
4 Push the O-ring to the front, and remove it.
4 Insert new O-rings into the recesses and spray them with KaVo Spray.

6.2.2 Changing the water filter

Hazard from contaminated products.
Infection hazard to the dentist and patient.
4 Prior to initial startup and after each use, reprocess the product and ac-
cessories.
Overheating of the tooth due to insufficient amount of cooling water.
Insufficient spray water can cause the medical device to overheat and damage
the pulp and tooth.
4 Check water filter and replace, if necessary.
4 Check the spray water channels and if necessary clean the spray nozzles
with the nozzle needle (Mat. no. 0.410.0921).
4 Unscrew the filter with the wrench (Mat. no. 1.002.0321) and take it out.
4 Insert a new filter (Mat. no. 1.002.0271) and screw it in with the wrench.
WARNING
CAUTION