Fuse Holder And Voltage Changer Assembly; Interconnections; Connection P1 Logic Input Interconnections - Varian Turbo-V 300 ICE Instruction Manual

Hide thumbs Also See for Turbo-V 300 ICE:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 77
TECHNICAL INFORMATION
FUSE HOLDER AND VOLTAGE CHANGER
ASSEMBLY
The following figure shows the location of this
assembly.
Rear panel
Proceed as follows to replace one or both fuses:
Remove the assembly by levering in position A with
a small screwdriver.
Replace the fuse
Use only T-type fuses of the following
characteristics:
- 100/120 Vac
- 220/240 Vac
Voltage Changer
To change the power supply voltage rating after having
removed the assembly as explained above, extract the
voltage changer and then reposition it to view the
desired voltage rating.
52
3.15 A
6.3 A

INTERCONNECTIONS

The following figure shows the Controller inter-
connections.
Controller models 969-9433
and 969-9533 interconnection
Connection P1
Logic Input Interconnections
P1 input connector
All the logic input to the controller must be connected at
J1 mating connector. With the provided J1 mating
connector (shipped with pin 3 and pin 8 shorted) make
the connections with AWG 24, (0.24 mm
wire to the pins indicated in the figure to obtain the
desired capability. The following table describes the
signals available on the connector.
TURBO-V 300 ICE CONTROLLER - 87-900-902-01 (G)
2
) or smaller

Advertisement

Table of Contents
loading

This manual is also suitable for:

969-9433969-9533

Table of Contents