CAUTION! Cautions are used when failure to observe instructions could result in damage to equipment, whether Varian supplied or other associated equipment. NOTE Infomation to aid the operator in obtaining the best performance from the equipment.
TECHNICAL INFORMATION DESCRIPTION CONTROLLER SPECIFICATIONS The Turbo-V 300 box controller is a microprocessor- Input: controlled, solid-state, frequency converter with self- - Voltage 75 Vdc diagnosis and protection features. with 2 Vpp max ripple - Current 3.0 A max. The controller drives the Turbo-V 300 pump series by controllling the voltage and current respect to the speed Fuse T 3 A...
TECHNICAL INFORMATION CONTROLLER OUTLINE The outline dimension for the controller are shown in the following figure: Controller outline 87-900-944-01 (C)
TECHNICAL INFORMATION INSTALLATION I/O Specifications Inspect the controller for any shipping damage. START/STOP: - START command Low <0.8 Vdc Should the controller be connected to a host computer via the-RS-232 interface, a suitable cable must be - STOP command High 4 to 15 Vdc prepared.
Message Structure (request and answer have the same format) NOTE The master system (PC) starts every session sending the following message to the slave units connected: Varian cannot guarantee compliance with regulations for radiated emissions unless all external wiring is shielded, with the shield being terminated to <STX>...
TECHNICAL INFORMATION When a slave device is addressed by the master: Command : SOFT-START (OFF) Source : PC 1) In case of reading request of the value contained in Destination : Inverter a window, the slave answers a string equal to the one sent by the master but in addition there is the field <DATA>...
TECHNICAL INFORMATION Serial Communication Windows TYPE Description X START/STOP (1= START ; 0= STOP) X REMOTE/SERIAL Configuration (1= Remote ; 0= Serial) X SOFT START YES/NO (1= YES ; 0= NO) Default= 0 X ACTIVE STOP (0=NO; 1=YES) Default= 1 X BAUD RATE (3-4) [4800-9600] Default= 4 X PUMP LIFE RESET [Write “1”...
Replacement controllers are available on an advance and refer to Turbo-V pump instruction manual prior to exchange basis through Varian. necessary, operating the Turbo-V controller. information is provided to aid the operator in determining malfunctions and corrective steps to be taken.
TECHNICAL INFORMATION PCB JUMPERS 87-900-944-01 (C)
Page 15
NOTE: If a product is received at Varian which is contaminated with a toxic or hazardous material that was not disclosed, the customer will be held responsible for all costs incurred to ensure the safe handling of the product, and is liable for any harm or injury to Varian employees as well as to any third party occurring as a result of exposure to toxic or hazardous materials present in the product.
Page 16
Request for Return FAILURE REPORT TURBO PUMPS and TURBOCONTROLLERS POSITION PARAMETERS Power: Rotational Speed: Does not start Noise Vertical Current: Inlet Pressure: Does not spin freely Vibrations Horizontal Temp 1: Foreline Pressure: Does not reach full speed Leak Upside-down Mechanical Contact Overtemperature Other: Temp 2:...
Page 17
Toll Free # 1 (800) 882 7426 Tel: 00 800 234 234 00 vtt.technical.support@varianinc.com China Mexico Varian Technologies - Beijing Varian, S. de R.L. de C.V. Rm 1648 Central Tower South Wing Concepcion Beistegui No 109 China Beijing Junefield Plaza Col Del Valle No.