TECHNICAL INFORMATION
CONTROLLER SPECIFICATIONS
Input:
Voltage
Frequency
Power
Output for pump:
Voltage
Frequency
Power
water cooling
air cooling
Operating temperature
Storage temperature
Fuse:
Mains
Heater Jacket (F4)
P1 optoisolator input
J2 optoisolator output
J7 output (fan, forepump coil, etc.)
J11 purge flow connector
J12 heater jacket connector
In compliance with norms
Auxiliary connectors
P1
J2
J7
J14, J13
Interconnecting cables
Weight (both models)
50
100, 120, 220, 240 Vac, 1-phase
50 to 60 Hz
650 VA maximum
80 Vac nominal
10%
933 Hz
2%
250 W maximum
250 W
150 W
0 °C to +40 °C
-20 °C to +70 °C
2 x T3.15A(slow blow) for 220 or 240 input Voltage
2 x T6.3A(slow blow) for 100 or 120 input Voltage
T3.15A (slow blow)
Minimum ON 3 mA
24 Vdc, 60 mA
120 Vac, 1 A disregarding the mains
12 Vdc power supply output
120 Vac 2A power supply output
EN 30011 class A group 1
EN 61010-1
IEC1000-4-2,1000-4-3, 1000-4-4
External INPUT signals (pins)
OUTPUT signals (sockets)
Output fan Voltage and Valve, forepump socket
RS 232/422/485 connection (optional)
Mains cable (3-wire, 3-meter long)
Pump cable (6-wire, 5-meter long)
11.3 kg (24.9 lbs)
TURBO-V 300 ICE CONTROLLER - 87-900-902-01 (G)
Maximum 5mA