Cleaning the Solvent Filters
• Remove the blocked solvent filter from the bottle-head assembly and place
• Thoroughly flush the filter with bidistilled water (remove all nitric acid,
• Replace the filter.
Never use the system without solvent filters. This could cause damage to the pump
C A U T I O N
valves

Inline filter

The nano pump is equipped with a filter in front of the EMPV.
The standard filter has a volume of typically 100 µl. If the application needs a
reduced volume (e.g. for fast gradient) the 20 µl low volume filter
(01090-68703) is recommended. Be aware that the filter efficiency and
capacity is significantly reduced compared to the standard one.
Never run the nano pump without an inline filter.
N O T E
The Nano LC/MS does not require a mixer, since baseline ripples as known
from UV detection are not issue for the MS.
1200 Series Nano Pump User Manual
it in a beaker with concentrated nitric acid (65%) for one hour.
some columns can be damaged by nitric acid).