Agilent Technologies 1200 Series User Manual page 26

Nano pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

2
Getting started
Setup pump (continued)
3 Solvents
Solvent composition
Solvent calibration
26
For each channel, you may select which of the two solvents
to deliver. You can set the percentage of Solvent B to any
value from 0 through 100%. Solvent A always delivers the
remaining volume, calculated as 100% - %B. The Solvent B
scrollbar allows you to turn Solvent B ON or OFF. The text
boxes allow you to type a brief description for each of the
solvents.
If you turn a solvent channel OFF, it will not be used in the
timetable.
If Micro Flow mode is selected, you may select a calibration
curve out of a set of predefined calibration tables that correct
the calibration for the composition of most standard solvent
combinations.
1200 Series Nano Pump User Manual