Volume flow, temperature and part quantity counter
The simulation starts as soon as the simulation is activated and the setting menu
is exited. The simulation is interrupted or stopped, if the setting menu is called up.
There are 3 different simulation types available for each purpose:
a.) „Triangle" mode
Simulation value
Increment value
Start value
process is repeated continuously until the set time "Sim Auto Stop Time" has
expired and the simulation ends.
b.) "Monotonic" mode
Simulation value
Increment value
Start value
c.) "Static" mode
Simulation value
Start
value
Simulation value
Inkrementwert
Counter state
before start
When simulating the subset counter, note that in triangle mode, the submenu
counter must be in bidirectional mode to achieve the expected effect.
A5E48762204-AA
Intervals
Time
Interval time
Sim Autostoptime
Intervals
Time
Interval time
Sim Autostoptime
In the "Static" mode, a constant value is output for
the measured values flow and temperature. The
simulation ends after the set simulation time.
Time
Sim Autostoptime
In the "static" simulation for the part quantity
counter, the counter is only changed once by the
set "increment value" after the start.
Time
Sim Autostoptime
SITRANS FM100
In "Triangle" mode, the simulation
value increases continuously in the
increment of the parameter
"Increment value" and in the interval
"Interval time" with the "Start value".
After the amount of the parameter
"intervals", the simulation value
decreases again in the same way,
in order to increase again. This
In "Monotonic" mode, the simulation
value increases continuously in the step
size of the parameter "Increment value"
and in the interval "Interval time" with the
"Start value". After the amount of
"Intervals" or the expiration of "Sim Auto
Stop Time", the simulation ends.