Summary of Contents for Stryker TPS MicroDrill 5100-15
Page 1
2/05 5100-15-700 Rev-E MicroDrill 5100-15 Use only Stryker attachments: • 5100-15-40 Straight Attachment • 5100-15-41 Long Straight Attachment • 5100-15-42 Angled Attachment • 5100-15-43 4:1 Straight Attachment US Patent: 6,752,816 Stryker Instruments European Authorized Rep: 4100 E. Milham RA/QA Manager...
Page 2
O.R. staff, delay surgery if not functioning opti- mally, or cause damage to handpiece. NEVER modify any accessory to fit the handpiece. Use only Stryker part number series: 277; 1607; 5300-10; and 5300-20. Use Stryker Irrigation Clips REF 5100-15-541 and...
Page 3
Instructions NOTE: The MicroDrill can be operated with the TPS Footswitch or the TPS Universal Handswitch. Assembly: WARNING: DO NOT attach any component while the motor is running. 1. Push the attachment onto the handpiece with a twisting motion. The attachment is secure when it "snaps"...
Page 4
Troubleshooting Problem Attachment difficult to attach or release from the handpiece. Cause Solution Connectors require Attach and detach the a breaking-in period. attachment repeatedly until connectors function smoothly. Build-up of debris See below. inside attachment. Problem Attachment does not retain bur. Cause Solution Lock collar not fully...