Suction
Port
Figure 3.15
Flushing Detergent Solution Into Air/Water Channels
Using the All-channel Irrigator (CW-3) and the 30 cm
the air and water channels according to the procedure described below.
1.
Attach the All-channel Irrigator's Channel cap to the Instrument Channel
Port (see Figure 3.16).
2.
Attach the All-channel Irrigator's Lock cover to the Air/Water and Suction
Cylinders (see Figure 3.16).
OSF-3 ENDOSCOPE REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Channel Cleaning
Adaptor
(MB-19)
Insertion Tube
Endoscope Connector
Suction Tank
The Suction System
The models such as the Channel Plug (MH-944) and the
Injection Tube (MH-946) cannot be used. Using such models
may result in instrument damage.
Instrument
Channel Port
Instrument
Channel
Universal Cord
Suction Connector
Suction Tube
Suction Pump
3
(30 ml) syringe, irrigate
Suction
Cylinder
Endoscope
31