Sliding Sleeves; Post-Mounting Check - Endress+Hauser Liquiphant FTL63 Density Operating Instructions Manual

Vibronic density measurement for liquids specifically for the food and life sciences industries
Hide thumbs Also See for Liquiphant FTL63 Density:
Table of Contents

Advertisement

Mounting
20
Closing the housing covers
NOTICE
Thread and housing cover damaged from dirt and fouling!
Remove dirt (e.g. sand) on the thread of the covers and housing.
If you continue to encounter resistance when closing the cover, check the thread again
for fouling.
Housing thread
The threads of the electronics and connection compartment can be coated with an
anti-friction coating.
The following applies for all housing materials:
Do not lubricate the housing threads.
5.3

Sliding sleeves

 For more details, see the "Accessories" section.
A
5.4

Post-mounting check

 Is the device undamaged (visual inspection)?
 Does the measuring device meet the measuring point specifications?
For example:
• Process temperature
• Process pressure
• Ambient temperature
• Measuring range
 Are the measuring point number and labeling correct (visual inspection)?
 Is the device adequately protected from wet conditions and direct sunlight?
 Is the device properly secured?