Examples of support in the event of dynamic load
5.1.7
Weld-in adapter with leakage hole
Weld in the weld-in adapter in such a way that the leakage hole is pointing downwards.
This enables any leaks to be detected quickly.
11
Weld-in adapter with leakage hole
5.2
Mounting the device
5.2.1
Required tool
• Open-ended wrench for sensor installation
• Allen key for housing locking screw
5.2.2
Installation
Aligning the tuning fork using the marking
The tuning fork can be aligned using the marking in such a way that the medium drains
off easily and buildup is avoided.
Markings are labeled on the process connection by means of:
Material specification, thread designation, circle, line or double line
Mounting
A0039742
A0039230
17