Avoid buildup

NOTICE
The presence of buildup or corrosion on the tuning fork affects the result of the
measurement and must be avoided!
Allow for maintenance intervals if required!
5.1.5

Take clearance into consideration

Allow sufficient space outside the tank for mounting, connection and replacement of the
electronic insert.
 9
Take clearance into consideration
5.1.6

Support the device

Support the device in the event of severe dynamic load. Maximum lateral loading capacity
of the pipe extensions and sensors: 75 Nm (55 lbf ft).
Liquiphant FTL63 Density
D
92 mm (3.62 in)
94 mm (3.70 in)
96 mm (3.78 in)
98 mm (3.86 in)
100 mm (3.94 in)
>100 mm (3.94 in)
r
1.0002
1.0001
1.0001
1.0001
1.0001
1.0000
A0039741