1. Safety instructions and warnings Read all instructions before using the machine. When using an electrical appliance, basic precautions should always be followed including the following. Caution! The operator of this instrument is advised that if the equip- tion provided by the equipment may be impaired. Warnings caused when the appliance is not used according to the and competent contractor.
Page 5
heating elements immediately after the end of a program as Dangerous voltage Do not hose the machine down to clean it. Do not sit or lean on the open door. This could cause the machine to tip and cause damage to the user and the Be aware of the steam discharge.
2. Intended use of the machine Items suitable for processing in a thermal disinfector cleaning and thermal disinfection of medical devices intended for re-use such as; disinfected in the DEKO 190 ward washer disinfector should have the following properties: Heat resistant to a temperature of up to 100 °C •...
3. Operating instructions 3.1. Washing and disinfection cycles 3.1.1 Standard cycles Flush Phase °C Time ~5 s ~5 s Approximate total time 1 min Short Phase °C Time ~5 s ~5 s Circulation wash with detergent 2 min Circulation rinse, hot water 15 s Disinfection 1 min...
Cycle parameters Recommended cycle phase values: Time Temperature Detergent Note min, sec [ t ] volume ml [Q] Circulation wash 1,00–20,00 0–99 Circulation rinse, hot water 0,15–5,00 Disinfection 0,01–5,00 80–95 Wait Note. If the cycle does not include disinfection phase, user •...
Detergents are powerful and must be handled with care. In case of accidental contact with skin or clothing rinse immediately with plenty of clean water.In case detergent gets in contact with eyes, seek immediate medical attention. Process chemicals Detergent: DEKO 190 ward washer disinfector is equipped with one Detergent requirements Use only liquid detergents.
4. Loading the machine Loading Opening and closing the door Manual door: Non-touch door: closed automatically. Bedpan bucket Start the ”NORMAL”-cycle. Bedpans are placed in the stationary turning stand in the door when the stand is in the upper position. Bedpans are emptied automatically when the door is closed.
Page 11
Urine bottles Start the “SHORT”-cycle. Kidney bowls Start the “SHORT”-cycle. Suction bottles washer. Start the ”INTENSIVE”-cycle. Shoes sure the shoes withstand a minimum clearance of 100 degrees. Start the “SHORT”-cycle.
Page 12
Working solution bottles Note! Do not pour soapy liquids or liquids which may cause Start the “INTENSIVE”-cycle. Instruments suggested when transporting the instruments to a service center from a longer distance. the instruments steady during the cycle to avoid any Start the “INTENSIVE”-cycle.
Page 13
Loading examples and accessories See items suitable for processing in a thermal disinfector on page 4.
Controls 4.2.1 Operating systems, manual door Main switch Door handle Panel printer (option) Detergent container Note. Touch screen display for user interface USER: User mode SERVICE: Service mode Alphanumeric keyboard to enter characters ENTER...
4.2.3 Running a cycle steam heated model, open also steam and condensate valves”. display appears and the machine is ready for use. (main menu) To next display (programs 4–8) To start To start To start To start SHORT -cycle NORMAL -cycle INTENSIVE -cycle If the machine is unused for a period of ten minutes it will go (manual door).
If the selected cycle does not include disinfection phase (e.g. 4.2.5 During a cycle the display will indicate 4.2.7 Malfunctions 4.2.6 End of cycle...
Warning text Explanation Loading door is open The loading door is open or not closed correctly. cycle. No disinfection disinfection phase. Cycle No. # OK, No disinfection disinfection phase. Open water valves Power cut during the cycle Will appear when machine requires preventive maintenance Periodic service is needed 4.2.9 Other menus...
15:48:33 Hot water rinse start 15:49:30 Hot water rinse end 15:49:33 Disinfection 90 °C start 15:55:26 Temperature attained T: 93 °C 15:56:26 Disinfection end T:91 °C 15:56:31 Disinfection time 60 s Cycle end time:06-08-21 15:56:45 Cycle passed ------------------------ Signature Rhima or it’s representatives.
5. Recycling and disposal instructions accordance with local regulations. Please beware of using and delivered to respective collection locations. appropriate safety wear (eg. and delivered to metal collection locations. gloves, respirator) when disassembling the machine. Please follow the instructions and regulations of the local It shall always be assumed that there may be harmful the instructions of the chemical’s supplier.
7. Declaration of conformity • Name and address of the manufacturer: • ity is declared: KWC Nordics Oy, Vartiokuja 1, 76850 Naarajärvi, FINLAND This declaration of conformity is issued under the sole EN 61010-1:2010 • • KWC Nordics Oy Safety requirements for electrical equipment for measure- DEKO 190, Ward Washer •...
Appendix 1: List of malfunctions Explanation Actions Door opened during the cycle. Open circuit in controlling temperature sensor T1. Short circuit in controlling temperature sensor T1. Open circuit in verifying temperature sensor T2. Short circuit in verifying temperature sensor T2. Temperature 15 °C over set value.
Appendix 2 User maintenance Purging and disinfecting the WD Methods of cleaning All parts of the machine supplying water and detergent to Daily: • • Weekly: • Clean the outer stainless steel panels with diluted washing • • facturer include a disinfection phase as the last phase of the Maintenance instructions process.
Page 24
detergent container and the date on which the machine is used. Appendix 2: User maintenance...
Page 25
• • • Wiping the outer casing and buttons 3. Flushing of chemical dosing systems Clean the stainless steel machine external surfaces and system regularly with hot water. parts of the machine are washed and disinfected during the detergent container is changed. Detergent tank replacement and dosing system rinse On machines where programs are run less frequently, and e.g.
Page 26
4. When necessary adjust the length of the suction pipe and place it in the new container Set the date and time in user’s mode Periodic deck maintenance...
Important Warranty Information service in commercial environments. TAKING DELIVERY OF YOUR RHIMA WASHER DISINFECTOR ‘TIP‘n’TELL indicates a fall has occurred during transportation. or caused whilst unloading and placement of the appliance. WARRANTY CONDITIONS AND COVERAGE a Service Agreement that includes a warranty extension component. In order to prevent potential premature failure of authorised service agent.
Page 28
Tel: 1300 347 944 New Zealand Tel: 0800 902 054 Singapore Tel: +65 9107 8943 DETERGENTS To request detergents or rinse additive contact your local Rhima Service centre below: Australia: 1300 347 944 New Zealand: 0800 902 054 Singapore: +65 9107 8943 Mediwash 5 Litre For washing glass &...