Maquet HANAULUX BLUE 130 Operating Instructions Manual page 21

Hide thumbs Also See for HANAULUX BLUE 130:
Table of Contents

Advertisement

6
Troubleshooting
Fault
Suspension / Mobility
Light is heavy to move
Light moves too easily
Lighthead falls/ rises
Optical system/ light mechanism
Light flickers
Light intensity too bright/ too dark
Light bulb doesn't last long enough Wrong light bulb(s)
Light field not round
Closure disc
Scratched or cracked closure disc
Sterilization
Sterilizable handle has too short
a life
Handles damaged or having
cracks
If a fault occurs repeatedly or cannot be remedied, please get in touch with your MAQUET contact partner.
56 351 001/F
Gebrauchsanweisung / Operating instructions HANAULUX
Possible cause
Brakes are set too tight
Brakes are set too loose
Spring force in weight
counterbalance / spring arm too
small/ too large
Contact problem
Wrong supply voltage
Wrong supply voltage
Wrong light bulb(s)
Light bulb seating incorrect
Main light bulb defective
Incorrect cleaning agent used,
Cleaning or disinfecting when hot
Wrong sterilization procedure
End of handles' life cycle
- 21 -
Remedy
Set brakes
Set brakes
Set spring force
Service
Service
Service
Check light bulb and if
necessary replace it
Check light bulb seating
Use different cleaning agent
Check sterilization procedure Sect. 3.2
Replace handles
®
blue 130/90
Reference
Sect. 5.3
Sect. 5.3
Sect. 5.2
---
---
---
Sect/ 5.1
Sect. 3.1