Performing Sweep Measurement; Stopping Measurement - Keysight Technologies B2961B Programming Manual

Low noise power source
Hide thumbs Also See for B2961B:
Table of Contents

Advertisement

Performing Sweep Measurement

Staircase sweep measurement is performed as follows.
1. Set the staircase sweep source and the required source functions. For details,
2. Set the required measurement functions. For details, see previous topics in this
3. Set the trigger condition. See
4. Enable the channel. See
5. Execute the :INIT command to start measurement.
For the programming example, see
To get measurement result data, use a :FETC subsystem command. For example,
the :FETC:CURR? command returns the latest current measurement data. The
:FETC?command returns the latest data for the parameters specified by the
:FORM:ELEM:SENS command.
For details on the :FETC subsystem commands, see SCPI Command Reference.

Stopping Measurement

Measurement is stopped by the :OUTP OFF command.
Example
ioObj.WriteString(":OUTP OFF")
Keysight B2961B/B2962B Programming Guide, Edition 1
see
"Controlling the Source Output" on page
section.
and
"Setting the Measurement Trigger" on page
The channel starts output set by the :SOUR:<CURR|VOLT> command.
Controlling the Keysight B2961B/B2962B
Controlling the Measurement Function
17.
"Setting the Source Output Trigger" on page 25
"Enabling the Measurement Channel" on page
"Staircase Sweep Output" on page
30.
81.
29.
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

B2962b

Table of Contents