Product Identification; Mounting; General; Explosion Hazardous Areas - Xylem MJK 713U Manual

Open channel flowmeter
Table of Contents

Advertisement

Manual

Product identification

General

A standard MJK 713 is delivered for two different measuring
ranges (0 - 3 ft and 0 - 10 ft).
It is very important that the measuring range is correct
for the flume in question in order to obtain maximum
measuring accuracy.
Check that the item(s) delivered corresponds to the or-
dered item(s). The item number is printed on a label that
is sticked onto the packing. Shown below is the label for
a MJK 713 System Package 1 delivery including a MJK
713 flowmeter and a Shuttle
Item number
Item description
An identical marking can
be found on the right
hand side of the level
transmitter cabinet:
US 3.10 713 FLOW CONVERTER 1905 - SW 833062
1.888.610.7664
ultrasonic sensor:
®
Serial number
Software version
www.calcert.com
15
10
20
5
25
0
30

Mounting

General
MJK 713 measures the flow in an open flume or weir
based on the level of the head in the flume. The method
of measurement are based on the following mathematical
function:
Flow Q = f(head
where the exponent x and the constant depends on the
type of weir or flume.
The flow converter has 3 different linearization
systems depending on how the volume of water is mea-
sured:
-
preprogrammed formulas for different dimensions of
the most common flumes and weirs, Parshall flumes,
Palmer & Bowlus flumes, V-notch weirs, and rectan-
gular weirs.
-
for other flumes and weirs with a known calculation
formula the exponent and the factor of the calcula-
tion formula can be set.
-
for flumes where no calculation formula exists, a
number of known flow values can be entered,
on which a point linearization and curve extrapolation
is calculated.
It is obvious that the flow calculation is highly dependant
on the accuracy of the level measurement. It is therefore
important to follow the mounting instructions carefully for
the Shuttle
Ultrasonic Sensor (MJK 713 System Package 1
®
) or MJK model 3100 Pressure Transmitter (MJK 713 System
Package 3).

Explosion hazardous areas

The Shuttle
ultrasonic sensor is Ex approved and can be
®
mounted in Zone 2 without the need of a
zener barrier. It is not approved for Class I Division I
zones.
The Flowmeter (= the electronic box with
display) must not be mounted in explosive
hazardous areas.
6
x
× constant)
sales@calcert.com