Mechanical mounting

MJK 713 Flowmeter

The MJK 713 Flowmeter is in NEMA 4X enclosure and
can be mounted outdoors directly on a wall or in a panel
using a panel mounting kit (order no. 200105). A sun
shield (order no. 200115) is recommended in very sunny
locations.
mounted vertically in order to observe the
Mechanical dimensions for the MJK 713 Flowmeter.
The drawing below shows the drilling distances for
mounting the cabinet on a wall:
A full scale print of above drawing can be found at the
end of this manual.
US 3.10 713 FLOW CONVERTER 1905 - SW 833062
1.888.610.7664
The MJK 713 Flowmeter must be
NEMA4X standard.
www.calcert.com
15
10
20
5
25
0
30
Shuttle
Ultrasonic Sensor
®
The following are extremely important when
mounting the ultrasonic transmitter:
1:
It should be mounted securely.
2:
It should be mounted absolutely vertical.
Use a spirit level in TWO directions.
3:
It should be mounted in the correct distance from
the bottom of the flume / weir.
Observe the blocking distance!
4:
It should be mounted in the correct distance from
the inlet.
1: The ultrasonic sensor is equipped with a nut for bracket
mounting. Note the recess on the nut - it must be fitted
safely in the bracket for firm fixing to the bracket:
2a: To ensure a reliable and precise level measurement it is
of vital importance that the ultrasonic sensor points down
absolutely vertical against the liquid surface.