Analyzer Initial Start-Up

4 ANALYZER INITIAL START-UP

PREPARING THE ANALYZER FOR START-UP

4.1
Once installation is complete, proceed as follows to prepare the analyzer for online operation :
1. Connect the sample line inlet tubing (or filtered sample outlet coming from optional
filtration system) to the fast-loop reservoir installed on the left side of the analyzer
2. Connect the drain fitting of the fast-loop reservoir to the waste line
3. Put the reagent inlet tubings into the appropriate reagent containers
a. Reagent 1 = Cresolphthalein Reagent
b. Reagent 2 = Ethanolamine Reagent
4. Put the calibration solution inlet tubing into the calibration solution positioned beneath
the analyzer
5. Connect the analyzer drain tubing to waste drain line
WARNING!
6. Check sample presence in fast-loop reservoir and adjust the sample flowrate so that there
is complete reservoir discharge when a finger is covering the hole at the top of the
stainless steel drain line
7. Turn the analyzer on using the switch in the electronics compartment
8. Perform reagents prime (see Section 6.1)
9. Perform calibration (see Section 6.2 & 6.3)
10. Set the analyzer to online mode (see Section 6.5)
29
Before proceeding with analyzer start-up it is absolutely
necessary to check that all procedures for a proper
installation and reagent preparation have been made.
Please verify that all the suggestions and recommendations
have been respected.
The drain of the analyzer must be at ambient pressure with
no restriction. Please verify that this condition has been
strictly respected during initial start-up procedures.
Waltron User Manual
3051 Hardness Analyzer