This equipment is intended for use in the electromagnetic environment specified below. The user of this
equipment should assure that is used in such an environment.
IEC 60601-1-2
IMMUNITY
TEST LEVEL
AGAINST
contact discharge: ± 6 kV
electrostatic discharge,
ESD (IEC 61000-4-2)
air discharge: ± 8 kV
power supply lines: ± 2 kV
electrical fast
transients / bursts
longer input / output lines: ± 1 kV
(IEC 61000-4-4)
surges on AC mains
Common mode: ± 2 kV
lines
differential mode: ± 1 kV
(IEC 61000-4-5)
power frequency
3 A/m
magnetic field 50/60
Hz
(IEC 61000-4-8)
voltage dips and short
dip >95%, 0.5 periods
interruptions on AC
dip 60%, 5 periods
mains input lines
dip 30%, 25 periods
(IEC 61000-4-11)
dip >95%, 5 seconds
Recommended separation distances between portable and mobile RF communications equipment
and the equipment
Separation distance according to frequency of transmitters in meters
RATED
MAXIMUM
OUTPUT POWER
OF TRANSMITTER
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
ELECTROMAGNETIC IMMUNITY
150 kHz – 80 MHz
d=1.2/V1]√P
0.12
0.38
1.2
3.8
12
866.4.Maxtec
COMPLIANCE
ELECTROMAGNETITC ENVIRONMENT
LEVEL (OF THIS
DEVICE)
± 6 kV
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
± 8 kV
relative humidity should be kept at levels to
reduce electrostatic charge to suitable levels.
± 2 kV
Mains power quality should be that of a typical
commercial or hospital environment.
± 1 kV
± 2 kV
Mains power quality should be that of a typical
commercial or hospital environment.
± 1 kV
3 A/m
Equipment which emits high levels of power
line magnetic fields (in excess of 3A/m) should
be kept at a distance to reduce the likelihood
of interference.
>95%, 0.5 per.
Mains power should be that of a typical commercial
or hospital environment. If user requires continued
60%, 5 per.
operation during power mains interruptions insure
30%, 25 per.
that batteries are installed and charged. Insure that
>95%, 5 sec.
battery life exceeds longest anticipated power
outages or provide and additional uninterruptible
power source.
80 MHz to 800MHz
d=1.2/V1]√P
0.12
0.38
1.2
3.8
12
www.maxtec.com
800 MHz to 2.5 GHz
d=2.3√P
0.23
0.73
2.3
7.3
23
25