NOTE: The operator must be facing the device and positioned within 4 meters to
distinguish the visual alarm indicators. Audible alarms can be distinguished as long as the
operator is in the same room and the ambient noise level is typical for a clinical setting.

5.0 CLEANING AND DISINFECTING THE MAXBLEND2

The external surfaces of the device and its accessories can be cleaned and
disinfected using the process detailed below. Under normal use conditions,
the sensing surfaces of the sensor should not become contaminated. If you
suspect that the sensing face of the sensor or internal surfaces of the flow
diverter have become contaminated, these items should be discarded and
replaced. Store the device in a clean, dry location when not in use.
1. Ensure battery drawer is closed and sensor/diverter are inserted into their port.
2. Spray CaviCide directly onto the surface to be cleaned.*
3. Wipe with clean paper towel to remove debris and bioburden.
4. Spray CaviCide directly onto pre-cleaned surface. Ensure that surfaces to be
disinfected are lightly but completely wetted.*
5. Apply additional spray as necessary to ensure the surfaces remain
continuously wet for 3 minutes.
6. Allow device to air dry completely.
7. Visually inspect the device for visible contamination. Repeat cleaning/disinfection
process if visible soil remains.
*NOTE: Only spray enough CaviCide to wet the surfaces. If the liquid begins to run,
drip or pool, remove it immediately using a dry paper towel and start over. Excess liquid
entering the device may cause internal damage.
DO NOT allow the liquid or spray to penetrate the device.
DO NOT spray cleaning solution directly onto the sensor port, bleed muffler
or buzzer openings.
Be sure to thoroughly clean and disinfect the areas depicted in the images
below. These regions are contacted during normal use and may contribute to
cross-contamination if not disinfected properly.
866.4.Maxtec