Maxtec MAXBlend 2 Operating Manual & Instructions For Use page 12

Hide thumbs Also See for MAXBlend 2:
Table of Contents

Advertisement

Low Battery Indicator -
7
middle of the display and is only activated when the voltage on the batteries
is below a normal operating level.
High Alarm Indicator -
8
below the "HIGH" icon on the LCD readout. The indicated value represents the
oxygen percentage at which the high alarm will be activated.
Calibration Reminder -
9
bottom of the display. This symbol will be lighted after one week has elapsed
from the previous calibration.
Low Alarm Indicator -
10
below the "LOW" icon on the LCD readout. The indicated value represents
the oxygen percentage at which the low alarm will be activated.
Up (Alarm High) Key -
11
limit. The device must be in the unlocked state for the key to operate. See
section 2.3.2 for instructions on setting the high FiO
Silent Key -
12
the audio alarm for 2 minutes.
ON/OFF Key -
13
device OFF, the button must be held while a rapid 3-2-1 countdown takes
place to prevent accidental power-off.
Calibration Key -
14
must be in the un-locked state for the key to operate. See section 2.8 for
instructions on calibrating.
Down (Alarm Low) Key -
15
alarm limit. The device must be in the unlocked state for the key to operate.
See section 2.3.1 for instructions on setting the low FiO
Un-lock Key -
16
Sleep Mode Indicator -
17
battery consumption.
Flow Control Knob - The flow control knob controls the flow of gas exiting
18
the flowmeter.
Oxygen Flowmeter - Measures the flow of mixed gas exiting the flowmeter outlet.
19
Backlight Key - The backlight key will manually activate the backlight for 30
20
seconds. See section 2.6 for more information on backlighting operation.
Smart Alarm Key - The Smart Alarm key is used to help set the High-Low Alarm
21
window quickly. See section 2.3.3 for instructions on using the Smart Alarm setting.
Bleed Toggle Switch - Turns on the auxiliary bleed. The bleed must be
22
turned on whenever the total flow delivered to the patient is less than 15
LPM for high flow model, or less than 3 LPM for low flow model.
4
The low battery indicator is located at the
BAT
The high alarm setting is displayed at all times just
The calibration reminder symbol is located at the
The low alarm setting is displayed at all times just
The up key is used in setting the high FiO
In an alarm condition, pressing the SILENT key will deactivate
This key is used to turn the device on or off. To turn the
This key is used to calibrate the device. The device
CAL
The down key is used in setting the low FiO
The un-lock key is used to unlock and lock the instrument.
The sleep mode indicator is used to help with
866.4.Maxtec
alarm limit.
2
www.maxtec.com
alarm
2
alarm limit.
2
2

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MAXBlend 2 and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

R229p01R229p02

Table of Contents