Instructions for Use
E N G L I S H
R230M06 REV. A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MaxO2 ME+p and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Maxtec MaxO2 ME+p

  • Page 1 MaxO ME+p ® Instructions for Use E N G L I S H R230M06 REV. A...
  • Page 2: Classification

    ME+p monitor to mable or explosive atmospheres may result in fire or explosion. be free from defects in material and workmanship for a period of two (2) years from Maxtec's date of shipment in a MaxO ME+p unit. Should a sensor fail prematurely, the replacement Never allow an excess length of cable near the patient’s head or neck, as such could result in...
  • Page 3 ◆ service technician or Maxtec for service. DO NOT disassemble the MaxO2 MaxO2 ME+p. All service should be performed by a Maxtec Certified Service Technician. DO NOT allow the sensor to come in contact with exhaled patient gases or other potential sources of contamination.
  • Page 4: Table Of Contents

    TABLE OF CONTENTS CLASSIFICATION ............2 3.5 External Power Supply Operation ..............11 WARRANTY .............2 4.0 SENSOR REMOVAL AND REPLACEMENT .. 11 WARNINGS ............2 5.0 PRESSURE MONITOR TESTS ......11 5.1 Pressure Monitor Accuracy Test ................ 11 SYMBOL GUIDE ............5 5.2 Pressure Monitor Functional Test ..............12 1.0 SYSTEM OVERVIEW.........5 6.0 TROUBLESHOOTING ........
  • Page 5: Symbol Guide

    A MAX-550E • Self-diagnostic check of analog and microprocessor circuitry. oxygen sensor outputs a voltage which is used by the MaxO2 ME+p to determine the concentration of • Battery level indicator.
  • Page 6: Essential Device Performance

    SMART ALARM KEY — The smart alarm key is used to help set the High-Low Alarm window quickly. This automatically sets oxygen alarms at ±3% or pressure alarms at ±2 cmH ENGLISH WWW.MAXTEC.COM • (800) 748-5355...
  • Page 7: Max-550E Oxygen Sensor

    WARNING: Electrical shock or damage to the equipment may occur if an inappropriate external power supply is used. Maxtec recommends using only the Maxtec approved external CAUTION: The flow diverter for the sensor is for use with flowing gases only.
  • Page 8: Adjusting Alarm And Calibration Modes

    BATTERY ICON : The battery bars indicate the approximate remaining power in the 50% x (990/1000) = 49.5%). Maxtec recommends that you re-calibrate after changing point- batteries. of-use elevation by more than 500 feet (150m). In addition, calibration is recommended if the user is unsure when the last calibration proce- dure was performed or if the measurement value displayed is in question.
  • Page 9: Zero Calibration

    HIGH icons will begin to flash indicating the SET OPERATING MODE. Select oxygen or pressure by pressing the power/mode key 2. Press the DOWN (LOW ALARM) key on the keypad. NOTE: The Low Alarm digits begin to flash indicating the Low Alarm manual setting. WWW.MAXTEC.COM • (800) 748-5355 ENGLISH...
  • Page 10: Pressure Monitoring And Alarms

    The filter protects the monitor from moisture and reduces the In this configuration, the MaxO2 ME+p will only monitor pressure in the upper portion of the risk of cross contamination. The pressure monitoring line and filter is single use and should...
  • Page 11: Alarm Conditions And Priorities

    3.5 External Power Supply Operation stopcock or tee To extend the life of the batteries a Maxtec approved external power supply can be purchased. Once connected to the unit, total power is supplied by the external power supply. The batteries are MaxO2 ME+p monitor still required to be in the unit and will provide emergency power in the event main AC power is lost.
  • Page 12: Pressure Monitor Functional Test

    (cm). Refer to diagram below. NOTE: Use only a Maxtec approved Max-550E sensor called out in Section 10.0 Spare Parts and 5. Submerse the end of the pressure monitoring line into the cylinder to the 5 cm Accessories.
  • Page 13: Alarm Testing

    The cable can be removed and replaced by disconnecting the threaded locking shroud at the Enflurane <1.5% sensor and monitor ends of the cable. Use only the Maxtec approved cable listed in Section 10.0 Spare Parts and Accessories. Isoflurane <1.5% NOTE: Ensure the cable locking shroud is fully threaded on the sensor and the monitor.
  • Page 14: Spare Parts And Accessories

    MaxO2 ME+p, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. WARNING: The MaxO2 ME+p should not be used adjacent to or stacked with other equip- ment. If adjacent or stacked use is necessary, the MaxO2 ME+p should be observed to verify normal operation.
  • Page 15 The MaxO2 ME+p was also tested for radiated immunity to RF wireless communication equip- ELECTROMAGNETIC IMMUNITY ment at the test levels below. IMMUNITY TEST COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT - GUIDANCE Frequency (Hz) Modulation Level V/m Electrostatic ± 8kV contact Floors should be wood, Pulse, 18 Hz, 50% DC concrete or ceramic tile.
  • Page 16 2305 South 1070 West Salt Lake City, Utah 84119 (800) 748-5355 www.maxtec.com...