Handi+N 2 O P E R A T I N G M A N U A L & I N S T R U C T I O N S F O R U S E R 2 1 8 M 0 6 I n d u s t r i a l R218M06 Rev.
Page 3
Product Disposal Instructions: The sensor, batteries, and circuit board are not suitable for regular trash disposal. Return sensor to Maxtec for proper disposal or dispose according to local guidelines. Follow local guidelines for disposal of other components.
Page 4
» The display is not valid when in Over Range Mode. Recalibrate the N2 and observe the proper operating procedure. » Never immerse the N2 or expose it to high humidity or moisture. It is not watertight. » Never expose the N2 to high temperatures. 866.4.Maxtec 866.4.Maxtec www.maxtec.com www.maxtecinc.com...
Do not throw away. Follow local Calibration Button guidelines for disposal. Contains acid Follow instructions for use Contains lead Do Not Meets ETL standards Serial Number Catalog Number Warnings Lot code/Batch code On/off Button IPX1 Drip Proof Manufacturer 866.4.Maxtec 866.4.Maxtec www.maxtec.com www.maxtecinc.com...
Oxygen Sensor: This is used to measure oxygen concentration in sample gas. Sample Inlet Connection: This is the port at which the device is connected to determine nitrogen concentration. 866.4.Maxtec 866.4.Maxtec www.maxtec.com www.maxtecinc.com...
The N2 analyzer will hold calibration and read correctly within ±3% when in thermal equilibrium within the operating temperature range. The device must be thermally stable when calibrated and allowed to thermally stabilize after experiencing temperature changes before readings are accurate. 866.4.Maxtec 866.4.Maxtec www.maxtec.com www.maxtecinc.com...
Page 9
For this reason, the following is recommended: Avoid usage in environments greater than 95% relative humidity. • 866.4.Maxtec 866.4.Maxtec www.maxtec.com www.maxtecinc.com...
Press and hold the Calibration Button for three seconds to manually force a new calibration. If unit repeats this error more than three times, contact Maxtec Customer Service for possible sensor replacement. CAL Err hi: Sensor voltage too high Press and hold the Calibration Button for three seconds to manually force a new calibration.
Operating Pressure: ......Atmospheric pressure to 3 psig Expected Storage Life : ....Two months with freshness seal on sensor. 866.4.Maxtec 866.4.Maxtec www.maxtec.com www.maxtecinc.com...