Electrical connection
7.4

TEDS transducer identification

TEDS (Transducer Electronic Data Sheet) allows you to
store the sensor characteristic values in a chip as per
IEEE1451.4. The C9C can be supplied with TEDS, which
is then mounted in the connector housing, connected and
supplied with data by HBM before delivery. If the force
transducer is ordered with a manufacturing certificate,
the characteristic values from the manufacturing
certificate are stored in the TEDS chip; if an additional
DAkkS calibration is ordered, the calibration results are
also stored in the TEDS chip.
The TEDS module is implemented in a Zero­Wire
configuration. The connector wiring is then such that the
force transducer can be connected to the HBM amplifier
with a Zero-Wire configuration. Please note that for
TEDS to function correctly, all extensions must be
implemented in a six wire configuration.
If a suitable amplifier is connected (e.g. QuantumX from
HBM), the amplifier electronics will read the TEDS chip
and parameterization will then be implemented
automatically, without any intervention required by the
user.
The chip content can be edited and configured with
suitable hardware and software. This can be
implemented, e.g. with the Quantum Assistant or even
the DAQ software CATMAN from HBM. Please follow the
operating manuals of these products.
C9C
23
A3729-2.2 HBM: public