General Installation Guidelines - HBM C9C Mounting Instructions

Hide thumbs Also See for C9C:
Table of Contents

Advertisement

Available languages

Available languages

F
Force in direction of measurement
nom
F
Force in direction of measurement, outside the
ex
center of the force transducer
e
Eccentricity (center of force transducer - force
application distance)
C9C
6.2

General installation guidelines

The forces to be measured must act on the transducer
as accurately as possible in the direction of
measurement. Bending moments resulting from lateral
force, eccentric loading and the lateral forces
themselves, may produce measurement errors and
destroy the transducer, if limit values are exceeded.
Fig. 6.1
Parasitic forces and moments
Notice
When installing and operating the transducer, please
note the maximum parasitic forces - lateral forces (due to
oblique application), bending moments (due to eccentric
force application) and torques, see the Specifications
section 9 on page 26, and the maximum permissible
loading capacity of the force application parts used
(possibly provided by the customer).
A3729-2.2 HBM: public
Mechanical installation
e
F
nom
Fq
F
ex
17

Advertisement

Table of Contents
loading

Table of Contents