Thermo Scientific Isotemp -86 C Freezers Assembly, Installation And Operation page 37

Table of Contents

Advertisement

S.
Function
Parameter
No
Code
Relay
25
Enclosure
0x03
Status
Isotemp -86°C Freezers
Address
Modbus
in hexa
Size
Command
decimal
3A 30 31
30 33 30 36
674
2
37 34 30 30
30 32 38 30
0D 0A
Relay
Data
Enclosure
Data
Type
Response
3A 31 39
30 33 30 34
0x00
30 30 30 35
0500
uint
30 30 30 30
00
44 37 0D
0A
Appendix B: Modbus ASCII Parameter Table |
Remarks
b0 - Bus Solenoid
Injection
b1 - Bus Pressure
switch
b2 - Reserved for
factory use only
b3 - Line voltage
circuit state change
(normal, buck, boost)
b4 - Compensated
line voltage change
b5 - Reserved for
factory use only
b6 - Short cycle
active
b7 - 4-20mA digital to
analog converter data
corrupt
b8 - Next Power up
state
b9 - Door1 Open
b10 - Door2 Open
b11 - Warm
temperature alarm
test Active
b12 - Read Reset
status register
b13 - Water cool
system pressure
sensor state (water
cooled units only)
b14 - 4-20mA digital
to analog converter
over temp fault
b15 - 4-20mA digital
to analog converter
over current of the
integrated chip fault
b16 - Main memory
corrupt
b17 - Back up
memory corrupt
33